TSHD - Cloned Gyro

Cloned Gyro System Definition

The underlying design of QINSy does not allow, for example, using different gyros and motion reference units for different computations on the same object. Nor is it possible to use a gyro system defined on one object to orient a second object when the second object is restrained to have the same heading as the first object.

As a remedy the System Cloner driver is used to duplicate certain system types without the need to physically split their interface channel (e.g. serial port). Essentially the driver copies the observation information from the original driver and makes it available to the rest of the system as if it was a separate observation.

In the case of a trailing suction hopper dredger, it may be permissible to assume the heading of the upper and lower suction pipes and/or the dredge head is the same as the vessel itself, though in many cases this is not a correct assumption. 

This section describes how a gyro heading is cloned in the event that a dredge head and/or lower pipe are deemed to have the same heading as the vessel. The upper pipe will rarely have the same heading.

Refer to TSHD - Positioning Dredge Heads for more information on the positioning of the trailing arm sections and the dredge head.

image2016-1-25 17:30:39.png In the tree right click on the name System and select New System. Alternatively click on the File menu, select New and then New System.

TSHD_DbSetup_ClonedGyro_NewSys.jpg

The opening dialog of the System wizard looks like this:

TSHD_DbSetup_ClonedGyro_NewSys_P1.jpg

image2016-1-25 17:37:0.png Enter a name for the cloned gyro system that you will easily recognize Online, differentiating it from the original gyro name.

image2016-1-25 17:38:30.png Select the cloned system sensor type. This must be the same as the original primary system type.

image2016-1-25 17:39:30.png Select the System Cloner driver.

DBSetup_EditSystem_ClonedGyro_DriverList.jpg

image2016-1-25 17:42:6.png Enter a unique number for each cloned system.

It does not matter what the number is as long as it is unique. If there are multiple cloned gyros, use a different port number for each one.

image2016-1-25 17:44:14.png Select the sensor system from which the cloned system will take its observation values.

All the sensor systems defined to date in the template database are listed. In the example these are:

TSHD_DbSetup_ClonedGyro_SysList.jpg

image2016-1-25 17:47:28.png The observation values copied from the primary sensor system can be modified by using an offset value.

Select the system that will provide the offset value. This system must be of the same type.
All the sensor systems defined to date in the template database are listed. In the example these are:

TSHD_DbSetup_ClonedGyro_OffsetSysList.jpg

For a gyro, the offset system might be a gyro with manual input. Such a system allows for a constant rotation of object 2 in relation to object 1.

image2016-1-25 17:52:18.png Click Next. The second page of the wizard looks like this:

TSHD_DbSetup_ClonedGyro_NewSys_P2.jpg

image2016-1-25 17:58:48.png Enter an easily recognized name for the cloned gyro observation when Online.

image2016-1-25 18:0:17.png Select the object upon which the cloned system is located.

image2016-1-25 18:1:34.png Enter all the observation parameters.

Observation Parameters

Type

Select Bearing (True), only option.

Unit

Select the type of units the gyro is outputting. This should be the same as for the primary gyro located on the pontoon.

A-priori SD

Enter the a-priori SD value of the gyro. Usually the default value is acceptable.

Fixed C-O

Enter fixed C-O correction. Fixed (C-O) corrections do not vary over time or location and must be recorded in the same measurement unit as the observation they refer to.

Variable C-O

Enter Variable C-O. The value for the Calculated minus Observed correction you enter will be added to the raw observation value.
The C-O correction is related to systematic minor deviations of the measurements. Variable (C-O) corrections are determined by calibration.
They may be instrument specific and/or time/location dependent.

Scale factor

The Scale Factor or Calculated divided by Observed correction should correct the raw observation measurement unit to the standard measurement unit.
When no Scale factor is needed, leave it at 1.0

image2016-1-25 18:2:44.png Enter the first slot number making it identical to the slot number used by the primary observation.

In this case no slot number is required as the original gyro observation for the TSHD had none. So these fields can be left blank.

image2016-1-25 18:7:21.png Click Finish to complete the system definition.

Return to top of page.

Return to: Trailing Suction Hopper Dredger (TSHD) - System Definitions