laser_zf_error_codes

Overview Error Codes for LEICA HDSxxx or Zoller+Fröhlich IMAGER/PROFILER Laser Scanner:

Firmware / SDK version: 8.9.0

Error Nr.

Error

Source

Description

0

ok

successfull

1

undef

undefined error

100

nio

Common error not further defined

101

sim

development

103

command param missing

command

Please read online manual for more information about command and params

104

command param wrong

command

Please read online manual for more information about command and params

105

Password wrong

user

106

Version conflict

firmware

107

command not implemented

command

The command you have tried is not defined

109

command param missing

command

Please read online manual for more information about command and params

110

no rights

command

You have not the rights to execute the command. You need superuser rights.

111

function disabled

command

112

currently not possible

progress

113

error open sio

hardware

115

in proccess, please wait

progress

116

user abort

user

The user cancelled the action

117

unknown fileformat

configuration

118

not available in this configuration

configuration

119

not possible at this time

progress

120

login failure

user

121

other user already logged in

user

122

logout must be loginuser

user

123

Pin not set

user

124

Pin wrong

user

132

socket connection to slow

scanning

200

USB read/write failed

hardware

202

USB chip FX2 not found

hardware

203

Upload program to FX2 failed

hardware

204

FX2 ASUSB

laserhardware

205

FX2 PWRCTRL

powerboard

206

FPGA file not found

laserhardware

220

FPGA upload failed

laserhardware

221

FPGA check no success

laserhardware

240

no connection to as over sio

laserhardware

241

power switched off

powerboard

400

no lara ini file

configuration

401

configfile write

configuration

402

no lara lut file

configuration

404

filter

configuration

405

wrong config entry

configuration

406

miss config entry

configuration

407

does not exist

configuration

409

problem load sw-lut

configuration

500

missing environment

firmware

603

set laser filter

laserhardware

621

Sensor too low

general

622

Sensor too high

general

623

Sensor absolut max

general

625

Temperaturecalib out of range

configuration

626

Undefined temperature calibration

configuration

627

Laserpower cannot be set

laserhardware

628

Filter can not be set

laserhardware

629

as reports error

laserhardware

634

Laser overwrite active

hardware

635

Problem warmup system

hardware

703

Battery empty

scanning

704

No battery pluged on

scanning

705

Powersupply voltage to low

scanning

800

Transmition error

scanning

801

Lost blocks

scanning

Some lines are lost while scanning.

802

Scan param

user

804

Start scan stream

scanning

810

harddisk access

scanning

811

Create zfs file

scanning

812

Disk full

scanning

Please use other disk

813

Timeout trigger

hardware

Scanner does not found start of scan. Please create report and call service

814

Timeout no data

hardware

Please retart otherwise create report and call service

816

Can't create folder

scanning

817

External disk access

scanning

818

FIFO overrun

hardware

Internal hardware problem

819

Laser monitoring

laserhardware

820

External disk no access, use intern disk instead

scanning

821

External disk no access

scanning

822

No usbstick found

user

823

File not found

user

824

Copy file failed

user

825

Disk run out of space

scanning

826

External disk access, use other external disk

scanning

827

Lasersafetysequence failed. Restart system.

hardware

Please restart system

828

Error on create next part of scanfile

scanning

829

External disk removed while scanning

scanning

830

Reference out of range

systemcheck

831

File already exist! Overwrite ?

user

832

No calibration for this settings. Calibration can be attached afterwards as well.

configuration

834

Disk full

scanning

Cleanup disk. Remove scans from disk

839

Low speed motor timeout

motor

840

process already running

user

841

no pause of previous scan

user

842

camera or motor not ready

camera

843

highest resolution without any selection

user

844

referenz intensity not between limits

laserhardware

845

modulator out of limits

laserhardware

846

emergency switch actice

hardware

Please check emergency switch is released

847

Referencmeasurement fault

hardware

Please call service

848

Referencmeasurement fault

hardware

Please call service

849

Referencmeasurement to less measurements

hardware

Please call service

900

Vert-Motor init

motor

No access to motor controller. May motor controller or powersupply demaged.

901

Vert-Motor referenz

motor

Problem on referenz

902

Vert-Motor speed

motor

Motor does not reached desired speed. Motor or mechanical problem.

903

Vert-Motor goto position

motor

905

Vert-Motor stop timeout

motor

906

Vert-Motor fault

motor

907

Vert-Motor speed. Please use warmup function

motor

951

Horz-Motor referenz

motor

952

Horz-Motor speed

motor

953

Horz-Motor goto position

motor

955

Horz-Motor stop timeout

motor

956

Horz-Motor fault

motor

957

Horz-Motor hold fault

motor

960

Problem configurate motor

motor

961

Configuration motor not ok

motor

963

Horz-Motor encoder index problem detected

motor

1000

Camera no access

camera

Camera is not detected. Please check cable, battery and camera is on.

1001

Camera not ready

camera

1003

Can't take image on camera

camera

1004

Can't detect camera

camera

1005

Can't save camera image

camera

1006

Error on calculate white-balance

camera

1008

Error copy white-balance file(s

camera

1010

Get global params failed

camera

1011

Graycard not recognised or not correct mounted

camera

Please check that graycard is on correct position

1012

camera light selection maybe wrong

camera

1013

i-cam selected but m-cam detected

camera

1014

problem getting cameraposition

camera

try one more

1050

problem accessing infrared camera

infrared

try one more

1051

problem getting image from infrared camera

infrared

try one more

1100

level not found

level

1102

level get value

level

1103

level out of range

level

1104

level moving while scanning detected

level

Scanner has moved while scanning. Detected by first and last measurement.

1105

not leveled

level

1106

level register failure

level

1120

dynamic compensator too fast movement detected. The affected lines will be filtered out later.

level

Area with to moch movement will clipped out. Or try additional scan

1121

dynamic compensator out of allowed range

level

to much movement or not leveled. Maybe try another scan

1122

dynamic compensator has no valid values

level

1206

Safety logic does not work

systemcheck

1208

One or more checks are disabled

systemcheck

1209

AS looked

laserhardware

Please call service

1210

Cover not removed

user

Please remove cover and try again

1301

gps receiver not found

gps

Could not find a GPS. GPS does not send NMEA strings. Maybe portsettings wrong.

1302

gps position too old

gps

GPS position is older than time defined in configuration

1303

gps signal too bad

gps

Too less satellits for accurate position. See configuration for limit.

1304

gps no position

gps

1305

gps no utc time

gps

1306

pps miss signal

gps

No PPS detected on counter1 input. Please check cable and signals.

1307

pps signal lost

gps

PPS signal lost while scanning. Please check GPS configuration, cable and signals.

1600

cameramotor

cameramotor

1601

cameramotor connect

cameramotor

Cameramotor of m-cam not found. Maybe not connected.

1602

cameramotor referenz

cameramotor

1603

cameramotor goto

cameramotor

1604

cameramotor timeout command

cameramotor

1606

cameramotor cmd wrong

cameramotor

1701

patch config unsuccessful

configuration

1702

wrong serial number

configuration

1703

config value not defined

configuration

2001

connect to zfnetstorage-server not possible, maybe not run or wrong settings

netstorage

2002

lost connection to zfnetstorage-server

netstorage

2004

watchdog timeout

profiler

2100

m-cam/i-cam

m-cam

2101

m-cam/i-cam not found

m-cam

m-cam not connected to the system

2102

m-cam(i-cam

m-cam

Problem initialise m-cam

2103

m-cam/i-cam setup

m-cam

2104

m-cam/i-cam capture

m-cam

Not possible to capture image

2105

m-cam/i-cam removed

m-cam

2150

i-cam

i-cam

2202

join slave

special

2500

Marker mode not supported

profiler

2501

Wrong param marker mode

profiler

2800

create image

calculation

2801

create preview image

calculation

2900

reference motor

p9000

3000

imu problem

imu

3001

imu problem

imu config miss

3002

imu not connected

imu

3003

no data from imu

imu

3004

imu hardware

imu

3005

imu command failed

imu

3100

prepaid account empty.

hardware

3101

prepaid recharge failed. Code wrong

hardware

3102

prepaid recharge failed. Code already used

hardware

3103

prepaid recharge failed. File not found

hardware

3200

faild to open scan

scan

3201

index not in scan

scan

3300

sensorboard not found

nav

3301

sensorboard version missmatch

nav